There appears to be a defect in the OpenLab version of the Agilent EZChrom Chromatography Data System that prevents results from being transferred to EZReporter automatically after acquisition. However, when you reanalyze the data file manually after the run, the results are transferred as expected. The defect prevents the “After Export” program, “ProcessEZChromData.exe”, from being called after acquisition.
To address this issue, we have written a new utility program called “EZ Result Monitor” that eliminates the need for the “After Export” program in the EZChrom method. It is a standalone program that watches the EZChrom export folder for the .ESD text files, and then processes the results contained in those files automatically as they are detected.
We have released a new version of EZReporter that contains this utility. You can download EZReporter 3.0.1.2 at this link:
https://diabloanalytical.com/ezr/files/ezreporter.3.0.1.2.sr1.exe
Instructions on configuring and using this utility can be found in the EZ Result Monitor reference manual:
https://diabloanalytical.com/ezr/files/EZ Result Monitor.pdf